- Smart interactive touchscreen control
- Newly optimized injection ports
- High stability and durability
Gas Chromatograph, GC-900A
Beschreibung
Description
The GC-900A gas chromatograph significantly enhances the accuracy and reproducibility of sample analysis through a meticulously design and optimized sample injection system.
Its circuit system has undergone a comprehensive upgrade, which not only substantially accelerates computation speed and improves data processing capabilities but also greatly enhances the instrument’s stability and durability. This creates a more robust and efficient analysis platform for users. As a result, the instrument can maintain long-term stable operation in various application environments, easily handling the diverse daily testing needs of users.
Features
Compatible Products & Accessories
Specification
| GC-900A | |||
| Host | Pressure/Flow Control | Mechanical control | |
| Number of sample ports supported | 3 | ||
| Maximum number of detectors supported | 4 | ||
| Temperature Control System | 8-channel independent temperature control system | ||
| Column oven | Control temperature | Room temperature+5℃-450℃ | |
| Minimum step value | 0.01℃ | ||
| Maximum heating rate | 120℃/min | ||
| Cooling time | From 450°C to 50°C in <5 minutes | ||
| Programmed Temperature Ramp | ≥20 steps | ||
| Temperature Stability | Ambient temperature ±1°C, column oven temperature variation <0.01°C | ||
| Others | 1. Automatic Furnace Temperature Protection Function | ||
| Injection System | Injection type | Compatible with split/splitless capillary injectors, septum purging packed column injectors, cold column head injectors, programmable temperature vaporization injection ports, valve injection systems, headspace injection systems, thermal desorption injection systems, etc. | |
| Maximum operating temperature | 450℃ | ||
| Programmatic current increase | ≥ 20 steps | ||
| Others | Features an independent split cold trap and a septum purge trap | ||
| Detector | Hydrogen Flame Ionization Detector (FID) | Maximum operating temperature | 450℃ |
| Detection limit | ≤4.3 pg C/s ( n-hexadecane ) | ||
| Baseline drift (30 min) | ≤2×10⁻13A | ||
| Baseline noise | ≤2×10⁻14A | ||
| Dynamic Linear Range | ≥10⁷ | ||
| Function | Auto-ignition function | ||
| Thermal Conductivity Detector (TCD) | Sensitivity | ≥8000mV.mL/mg | |
| Baseline drift (30 min) | ≤0.2mV | ||
| Baseline noise | ≤0.05mV | ||
| Dynamic Linear Range | ≥105 | ||
| Function | Air Gap Flow Protection Function | ||
| Category | Title | File type/size | Download |
| Brochure | Brochure-Gas Chromatograph | PDF(2 MB) | download |